Commands listed in alphabetical order
MEASUrement:
REFLevels:PERCent:HYSTeresis
This command sets or queries the p ercentage (where 100% is equal to MAX and
0% is equal to MIN) used to calculate the hysteresis of the reference level when
the measurem
ent's ref level m ethod is set to percent.
Group
Measurement
Syntax
MEASUrement:REFLevels:PER Cent:HYSTeresis <NR3>
MEASUrement:REFLevels:PER Cent:HYSTeresis?
Arguments
<NR3> is the percentage used to calculate the hysteresis of the reference level.
Examples
MEASUrement:REFLevels:PER Cent:HYSTeresis 3 sets the hysteresis to 3%.
MEASUrement:REFLevels:PER Cent:HYSTeresis? might return
:MEASUREMENT:REFLEVELS:PE RCENT:HYSTERESIS 5.0000 indicating the
hysteresis is set to 5%.
MEASUrement:REFLevels:PERCent:RISEHigh
This command sets or queries the percentage (where 100% is equal to TOP and
0% is equal to BASE) used to calculate the high reference level of the rising edge
when the measurement's ref level method is set to percent.
Group
Measurement
Syntax
MEASUrement:REFLevels:PER Cent:RISEHigh <NR3>
MEASUrement:REFLevels:PER Cent:RISEHigh?
Arguments
<NR3> is the percentage used to calculate the high reference level of the rising
edge.
Examples
MEASUrement:REFLevels:PER Cent:RISEHigh 90 sets the h igh reference
level to 90%.
MEASUrement:REFLevels:PER Cent:RISEHigh? might return
:MEASUREME
NT:REFLEVELS:PERCENT:RISEHIG H 80.0000
indicating the
high ref level is 80%.
2-504 MSO54, MSO56, MSO58, MSO58LP Programmer