Commands listed in alphabetical order
Examples
HORizontal:SA
MPLERate:ANALYZemode:MINimum :VALue AUTOMATIC
allows the oscilloscope to set the minimum value.
HORizontal:S
AMPLERate:ANALYZemode:MINi mum:VALue?
might return
:HORIZONTAL:SAMPLERATE:AN ALYZEMODE:MINIMUM:VALUE A UTOMATIC
indicating the oscilloscope automatically sets the minimum value.
HORizontal
:SCAle
This command sets or queries the horizontal scale.
Group
Horizontal
Syntax
HORizontal:SCAle <NR3>
HORizontal:SCAle?
Arguments
<NR3> is the horizontal scale in time per division..
Returns
The current horizontal scale is returned.
Examples
HORIZONTAL:SCALE 20e-9 sets the horizontal scale to 20 ns/division.
HORIZONTAL:SCALE? might return :HORIAONTAL:SCALE 20e-9 indicating the
horizontal scale is set to 20 ns/division.
ID? (Query Only)
This query-only command returns identifying information about the instrument
and related firmware similar to that returned by the *IDN? IEEE488.2 common
query but does not include the instrument serial number.
Group
Miscellaneous
Syntax
ID?
Related Commands
*IDN?
2-356 MSO54, MSO56, MSO58, MSO58LP Programmer