Commands listed in alphabetical order
POWer:POWer<
x>:SOA:VSOURce
This command sets or queries the voltage source for SOA measurement in the
specified power measurement badge. The power measurement badge is specified
by x.
Conditions
R
equires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:SOA:VSOURc e {CH<x> | MATH<x> | REF<x>}
POWer:POWer<x>:SOA:VSOURc e?
Arguments
CH<x> = A channel specifier; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x> = A math waveform specifier; <x> is ≥1.
REF<x> = A reference waveform specifier; <x> is ≥1.
Examples
POWer:POWer1:SOA:VSOURce CH1 sets the voltage source for SOA
measurement as channel 1.
POWer:POWer<x>:SWITCHINGLOSS:DEVICEType
This command sets or queries the conduction calculation method for switching
loss measurement in the specified power measurement badge. The power
measurement badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:SWITCHINGL OSS:DEVICEType {MOSFET | BJT}
POWer:POWer<x>:SWITCHINGL OSS:DEVICEType?
Examples
POWer:POWer1:SWITCHINGLOS S:DEVICEType MOSFET sets the conduction
calculation method as mosfet for switching loss measurement of the power
measurement badge Power 1.
2-586 MSO54, MSO56, MSO58, MSO58LP Programmer