Commands listed in alphabetical order
Examples
MEASUREMENT:C
LOCKRECOVERY:LOOPBANDWIDTH 2.0e6
sets the loop
bandwidth to 2.0 MHz.
MEASUREMENT:
CLOCKRECOVERY:LOOPBANDWIDT H
? might return
:MEASUREMENT:CLOCKRECOVERY :LOOPBANDWIDTH 1.0000E+6 indicating the
loop bandwidth is 1.0 MHz.
MEASUremen
t:CLOCKRecovery:MEANAUTOCalculate
This command sets or queries how often the c lock is calculated when constant
clock recovery is used for the measurement. This will affect measurements whose
:MEASUre
ment:MEAS<x>:CLOCKRecovery:GLOBal flag is set to 1.
Group
Measurement
Syntax
MEASUrement:CLOCKRecovery: MEANAUTOCalculate {FIRST| EVERY}
MEASUrement:CLOCKRecovery: MEANAUTOCalculate?
Arguments
FIRST calculates the clock on the first acquisition.
EVERY calculates the clock on every acquisition.
Examples
MEASUREMENT:CLOCKRECOVERY: MEANAUTOCALCULATE EVERY calculates the
clock on every acquisition.
MEASUREMENT:CLOCKRECOVERY: MEANAUTOCALCULATE? might return
:MEASUREMENT:CLOCKRECOVERY :MEANAUTOCALCULATE FIRST indicating the
clockiscalculatedonthefirst
acquisition.
MEASUrement:CLOCKRecovery:METHod
This command sets or queries the global clock recovery method
for the measurement. This will affect measurements whose
:MEASUrement:MEAS<x>:CLOCKRecovery:GLOBal flag is set to 1.
Group
Measurement
Syntax
MEASUrement:CLOCKRecovery: METHod
{PLL|CONSTANTCLOCK|EXPLICI TCLOCK}
MEASUrement:CLOCKRecovery: METHod?
MSO54, MSO56, MSO58, MSO58LP Programmer 2-397