Commands listed in alphabetical order
Examples
MEASUREMENT:C
LOCKRECOVERY:EXPLICITCLOCKMO DE EDGE
sets the clock
mode to edge.
MEASUREMENT:
CLOCKRECOVERY:EXPLICITCLOC KMODE
? might return
:MEASUREMENT:CLOCKRECOVER Y:EXPLICITCLOCKMODE EDGE indicating
the clock mode is edge.
MEASUremen
t:CLOCKRecovery:JTFBandwidth
This command sets or queries the global JTF bandwidth used when PLL clock
recovery is used for the measurement. This will affect measurements whose
:MEASUre
ment:MEAS<x>:CLOCKRecovery:GLOBal flag is set to 1.
Group
Measurement
Syntax
MEASUrement:CLOCKRecovery :JTFBandwidth <NR3>
MEASUrement:CLOCKRecovery :JTFBandwidth?
Arguments
<NR3> is the global clock recovery JTF bandwidth.
Examples
MEASUREMENT:CLOCKRECOVERY :JTFBANDWIDTH 2.0e6 sets the bandwidth
to 2.0 MHz.
MEASUREMENT:CLOCKRECOVERY :JTFBANDWIDTH? might return
:MEASUREMENT:CLOCKRECOVER Y:JTFBANDWIDTH 1.0000E+6 indicating the
bandwidth is 1.0 MHz.
MEASUrement:CLOCKRecovery:LOOPBandwidth
This command sets or queries the global loop bandwidth used when PLL clock
recovery is used for the measurement. This will affect measurements whose
:MEASUrement:MEAS<x>:CLOCKRecovery:GLOBal flag is set to 1.
Group
Measurement
Syntax
MEASUrement:CLOCKRecovery :LOOPBandwidth <NR3>
MEASUrement:CLOCKRecovery :LOOPBandwidth?
Arguments
<NR3> is the global loop bandwidth.
2-396 MSO54, MSO56, MSO58, MSO58LP Programmer