Commands listed in alphabetical order
MEASUrement:M
EAS1:REFLevels1:PERCent:TYPE ?
might return
:MEASUREMENT:MEAS1:REFLEVE LS1:PERCENT:TYPE TENNINET Y indicating
the reference levels percent type is 10%, 50% and 90% respective ly
MEASUrement
:MEAS<x>:REFMode
This command sets or queries the reference level mode for the measurement. The
measurement number is specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:REFMod e {AUTO|MANual}
MEASUrement:MEAS<x>:REFMod e?
Arguments
AUTO sets the reference level for the measurement automatically.
MANual allows the user to set the reference level for the measurement.
Examples
MEASUrement:MEAS1:REFMode Manual allows the user to set the reference
level for the measurement.
MEASUrement:MEAS1:REFMode? might return
:MEASUREMENT:MEAS1:REFMODE AUTO indicating the reference levels for the
measurement are set automatically.
MEASUrement:MEAS<x>:REFVoltage
This
command sets or queries the reference voltage value for the measurement.
The measure ment number is specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:REFVol tage <NR3>
MEASUrement:MEAS<x>:REFVol tage?
Arguments
<NR3> is the reference voltage value for the selected conguration.
Examples
MEASUrement:MEAS1:REFVolta ge 30.E-3 sets the reference voltage value
fo
r the measurement to 30 mV.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-479

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals