Commands listed in alphabetical order
Group
Power
Syntax
POWer:POWer<x>:REFLevels:P ERCent:HYSTeresis <NR1>
POWer:POWer<x>:REFLevels:P ERCent:HYSTeresis?
Related Commands
POWer:POWer<x>:REFLevels:METHod
Arguments
<NR1> ranges from 1 to 99.
Examples
POWer:POWer1:REFLevels:PER Cent:HYSTeresis 25 sets the hysteresis
as 25 percentage.
POWer:POWer<x>:REFLevels:PERCent:RISEHigh
This command sets or queries the rising edge for high reference level in
percentage for the specified power measurement badge. The power measurement
badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:REFLevels:P ERCent:RISEHigh <NR1>
POWer:POWer<x>:REFLevels:P ERCent:RISEHigh?
Related Commands
POWer:POWer<x>:REFLevels:METHod, POWer:POWer
<x>:REFLevels:
PERCent:TYPE
Arguments
<NR1> ranges from 1 to 99.
Examples
POWer:POWer1:REFLevels:PER Cent:RISEHigh 85 sets the high reference
level for rising edge as 85 percentage.
POWer:POWer<x>:REFLevels:PERCent:RISELow
This command sets or queries the rising edge for low reference level in pe rcentage
for the specified power measurement badge. The power measur ement badge
is specified by x.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-559