Commands listed in alphabetical order
Examples
WFMOUTPRE:YUN
IT?
might return :WFMO UTPRE:YUNIT "dB", indicating that
the vertical units for the waveform are measured in decibels.
WFMOutpre:YZEro? (Query Only)
This query-only command returns the combined vertical position and offset for the
source waveform specied by DATa:SOUrce. This represents a departure from
previous instruments where the :VZEro value represented the vertical position
in vertica
l units and the :VOFf v alue represented the vertical offset in digitizing
levels. For this instrument family, the value of :YOFf is always 0.0.
An error i
sreportediftheDATa:SOUrce waveform does not exist.
Group
Waveform Transfer
Syntax
WFMOutpre:YZEro?
Relate
d C ommands
DATa:SOUrce, WFMOutpre:YUNit?
Examples
WFMOUTPRE:YZERO? might return :WFMOUTPRE:YZERO -100.0000E-3,
indicating that vertical offset is set to -100 mV.
2-900 MSO54, MSO56, MSO58, MSO58LP Programmer