Commands listed in alphabetical order
Examples
DVM:SOURCE CH4
sets the DVM source to Channel 4.
DVM:TRIGger:FREQuency:COUNTer
This command s ets o r queries the state of the trigger frequency counter readout
in the trigger badge. This command requires the DVM option (free with product
registration).
Group
DVM
Syntax
DVM:TRIGger:FREQuency:COUN Ter {0|1|OFF|ON}
DVM:TRIGger:FREQuency:COUN Ter?
Arguments
1 or ON turns o n the trigger frequency counter for the Digital Voltmeter.
0 or OFF turns it off.
Examples
DVM:TRIGGER:FREQUENCY:COUN TER 0 turns off the counter.
DVM:TRIGGER:FREQUENCY:COUN TER? might r
eturn
:DVM:TRIGGER:FREQUENCY:COU NTER 1 indicating the counter is on.
*ESE
This command sets and queries the bits in the Event Status Enable Register
(ESER). The ESER prevents events from being reported to the Status Byte
Register (STB). For a more detailed discussion of the use of these registers, see
Registers.
Group
Status and Error
Syntax
*ESE <N R1>
*ESE?
Related Commands
*CLS, DESE, *ESR?, EVENT?, EVMsg?, *SRE, *STB?
Arguments
<NR1> specifies the binary bits of the ESER according to this value, which ranges
from 0 through 255.
The power-on default for the ESER is 0 if *PSC is 1. If *PSC is 0, the ESER
maintains the previous power cycle value through the current power cycle.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-321