Commands listed in alphabetical order
SEARCH:SEARCH
1:TRIGGER:A:BUS:MIL1553B:DAT A:PARITY?
might return
NOCARE, indicating that d ata words are not being filtered based on the parity
bit value.
SEARCH:SEAR
CH<x>:TRIGger:A:BUS:MIL1553B:DATa:VA Lue
This command sets or queries the value when searching on data words for a
MIL-STD-1553 bus. The search number is speci fi ed by x. The search condition
must be set
to DATA.
Conditions
Requires the SR-AERO Triggering and Analysis application.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A :BUS:MIL1553B:DATa:VALue <QString>
SEARCH:SEARCH<x>:TRIGger:A :BUS:MIL1553B:DATa:VALue?
Arguments
<QString> is the data value.
Examples
SEARCH:SEARCH1:TRIGGER:A:B US:MIL1553B:DATA:VALUE
"XXXXXXXXXXXX1000"
sets the value to XXXXXXXXXXXX1000.
SEARCH:SEARCH1:TRIGGER:A:B US:MIL1553B:DATA:VALUE? might
return "XXXXXXXXXXXXXXXX", indicating that the value is
XXXXXXXXXXXXXXXX.
SEARCH:SEARCH<x>:TRIGger:A:BUS:MIL1553B:ERRTYPe
This command sets or queries the type of error condition for a MIL-STD-1553
bus to search on. The search number is specified by x. The s earch condition
must be set to ERRor.
Conditions
Requires the SR-AERO Triggering and Analysis application.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A :BUS:MIL1553B:ERRTYPe
{PARity|SYNC|DATA}
SEARCH:SEARCH<x>:TRIGger:A :BUS:MIL1553B:ERRTYPe?
MSO54, MSO56, MSO58, MSO58LP Programmer 2-687