Commands listed in alphabetical order
MATH<x> =Amath
waveform specier; <x> is 1.
REF<x> = A reference waveform specier; <x> is 1.
Examples
POWer:POWer1:CYCAmp:INPUT SOurce CH1 sets the input source for cycle
amplitude measurement as CH1 for the power measurement badge Power 1.
POWer:POWer<x>:CYCLEBase:INPUTSOurce
This command sets or queries the input source for cycle base measurement in the
specied power measurement badge. The power measurement badge is specied
by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:CYCLEBase: INPUTSOurce {CH<x> | MATH<x> | REF<x>}
POWer:POWer<x>:CYCLEBase: INPUTSOurce?
Arguments
CH<x> = A channel specier; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x> = A math waveform specier; <x> is 1.
REF<x> = A reference waveform specier; <x> is 1.
Examples
POWer:POWer1:CYCLEBase:IN PUTSOurce CH2 sets the input source for cycle
base measurement as CH2 for the power measurement badge Power 1.
POWer:POWer<x>:CYCLEMAX:INPUTSOurce
This command sets or queries the input source for cycle maximum measurement
in the specied power measurement badge. The power measurement badge
is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
2-530 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals