Commands listed in alphabetical order
Group
Calibration
Syntax
CALibrate:PWRUpstatus?
Returns
This query will return one of the following:
0 Indicating the power-up calibration failed.
1 Indicating the power-up calibration passed.
Examples
CALIBRATE:PWRUPSTATUS? might return 0, indicating that the power-up
calibration failed.
CH<x>? (Query Only)
This query-only command returns the vertical parameters for the specified
channel. The channel is specified by x.
Group
Vertical
Syntax
CH<x>?
Examples
CH1? might return the following vertical parameters for
channel 1:
:CH1:BANDWIDTH 1.0000E+09;COUPLING
DC;DESKEW0.0000E+00;OFFSE T 0.0000E+00;POSITION
0.0000E+00;SCALE 5.0000E-01;PROBCAL INIT;PROBE:GAIN
1.0000E+00;RESISTANCE 1.0000E+06;UNITS "V";ID:TYPE
"1X";SERNUMBER "N/A";:CH1:PROBEFUNC :EXTATTEN
1.0000E+00;EXTUNITS "None";:CH1:LAB EL:NAME "";XPOS
5;YPOS 5
.
CH<x>:BANdwidth
This command sets or queries the selectable low-pass bandwidth limit filter of
the specified channel. The channel is specified by x.
The query form of this command always returns the approximate realized
bandwidth of the channel.
Available arguments depend upon the instrument and the attached accessories.
Group
Vertical
2-188 MSO54, MSO56, MSO58, MSO58LP Programmer