EasyManuals Logo
Home>Tektronix>Test Equipment>MSO58

Tektronix MSO58 User Manual

Tektronix MSO58
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #892 background imageLoading...
Page #892 background image
Commands listed in alphabetical order
TRIGger:A:HO
LDoff:TIMe
This command sets or queries the A trigger holdoff time. This command is
equivalent to selecting Mode & Holdoff from the Trig menu, selecting Time, and
then setting
the desired Holdoff Time.
Group
Trigger
Syntax
TRIGger:A:HOLDoff:TIMe <NR3>
TRIGger:A:HOLDoff:TIMe?
Related Commands
TRIGger:A:HOLDoff:BY
Arguments
<NR3> species the holdoff time in seconds. The range is from 0 seconds through
10 seconds.
Examples
TRIGGER:A:HOLDOFF:TIME 10 sets the A trigger holdoff time to 10 s.
TRIGGER:A:HOLDOFF:TIME? might return :T RIGGER:A:HOLDOFFTIME
1.2000E-06
, indicating that the A trigger holdoff time is set to 1.2 µs.
TRIGger:A:LOGICQUALication
This command sets or queries the type of logic qualication to perform.
Group
Trigger
Syntax
TRIGger:A:LOGICQUALificat ion {AND|OR|NAND|NOR}
TRIGger:A:LOGICQUALificat ion?
Arguments
AND species to trigger if a ll conditions are true.
NANd species to trigger if any of the c onditions are false.
NOR species to trigger if all conditions are false.
OR species to trigger if any of the conditions are true.
Examples
TRIGGER:A:LOGICQUALIFICAT ION OR species to trigger if any of the
conditions a
re true.
2-876 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals