Commands listed in alphabetical order
SEARCH:SEARC
H<x>:TRIGger:A:SETHold:LEVel:REF<x>
This command sets or queries the voltage level to use for setup & hold trigger
search. The search number is specifi ed by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:SETHold:LEVel:REF<x> <N R3>
SEARCH:SEARCH<x>:TRIGger: A:SETHold:LEVel:REF<x>?
Arguments
<NR3> is the voltage level to use for setup & hold trigger search.
Examples
SEARCH:SEARCH1:TRIGger:A: SETHold:LEVel:REF1 50.0e- 3 sets the
level to 50.0 mV.
SEARCH:SEARCH1:TRIGger:A: SETHold:LEVel:REF1? might return
:SEARCH:SEARCH1:TRIGGER:A :SETHOLD:LEVEL:REF1 0.0E+ 0 indicating
the level is 0.0 V.
SEARCH:SEARCH<x>:TRIGger:A:SETHold:LOGICPattern:CH<x>
This command sets or queries the conditions used for generating an A logic
pattern, with respect to the defined input pattern, and identifies the time that the
selected pattern may be true and still generate the trigger. The search number is
specified by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:SETHold:LOGICPattern:CH<x>
{INCLude|DONTInclude}
SEARCH:SEARCH<x>:TRIGger: A:SETHold:LOGICPattern:CH<x>?
Arguments
INCLude specifies including the specified channel SETHOLD inputs in the
specified search.
DONTInclude specifies not including the specified channel SETHOLD inputs
in the specified search.
Examples
SEARCH:SEARCH1:TRIGger:A: SETHold:LOGICPattern:CH1 INCLUDE
specifies including the specified channel SETHOLD inputs in the specified search.
2-734 MSO54, MSO56, MSO58, MSO58LP Programmer