Commands listed in alphabetical order
Syntax
POWer:POWer<x
>:REFLevels:ABSolute:TYPE
{SAME | UNIQue}
POWer:POWer<x>:REFLevels: ABSolute:TYPE?
Related Commands
POWer:POWer<x>:REFLevels:METHod
Arguments
SAME: select when the rising edge and falling edge reference levels are same.
UNIQue: select when the rising edge and falling edge reference levels are different.
Examples
POWer:POWer1:REFLevels:AB Solute:TYPE UNIQue setsthetypeof
measurement levels as unique for the specified power measurement badge.
POWer:POWer<x>:REFLevels:BASETop
This command sets or queries the reference level base top method for the specified
power measurement badge. The power measurement badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:REFLevels: BASETop {AUTO | MINMax | MEANhistogram
| MODEhistogram | EYEhistogram}
POWer:POWer<x>:REFLevels: BASETop?
Related Commands
POWer:POWer<x>:REFLevels:METHod
Examples
POWer:POWer1:REFLevels:BA SETop AUTO sets the reference level base top
method as auto for the power measurement badge 1.
POWer:POWer<x>:REFLevels:METHod
This command sets or queries the method to configure reference level values
for the specified power measurement badge. The power measurement badge
is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
2-556 MSO54, MSO56, MSO58, MSO58LP Programmer