Commands listed in alphabetical order
Group
Trigger
Syntax
TRIGger:{A|B}:PULSEWidth:W HEn
{LESSthan|MOREthan|EQual|U NEQual|WIThin|OUTside}
TRIGger:{A|B}:PULSEWidth:W HEn?
Related Commands
TRIGger:{A|B}:PULSEWidth:LOWLimit
TRIGger:{A|B}:PULSEWidth:HIGHLimit
TRIGger:{A|B}:PULSEWidth:SOUrce
Arguments
LESSthan causes a trigger when a pulse is detected with a width less than the
time set by the TRIGger:{A|B}:PULSEWidth:LOWLimit command.
MOREthan causes a trigger when a pulse is detected with a width greater than the
time set by the TRIGger:{A|B}:PULSEWidth:LOWLimit command.
EQual causes a trigger when a pulse is detected with a width equal to the time
period specified in TRIGger:{A|B}:PULSEWidth:LOWLimit within a ±5%
tol
erance.
UNEQual causes a trigger when a pulse is detected with a width
greater than or less than (but not equal) the time period specified in
TRIGger:{A|B}:PULSEWidth:LOWLimit within a ±5% tolerance.
WIThin causes a trigger when a pulse is detected that is within a range set by
two values.
OUTside causes a trigger when a pulse is detected that is outside of a range set
by two values.
Examples
:TRIGger:B:PULSEWidth:WHEn LESSthan causes the oscilloscope to
trigger when a pulse is detected that is shorter than the time s pecified by
TRIGger:{A|B}:PULSEWidth:LOWLimit.
TRIGger:A:PULSEWidth:WIDth
? might return
:TRIGGER:A:PULSEWIDTH:WHEN GREATER THAN 2.0000E-9 indicating that
a trigger is generated when a pulse is detected greater than 2 ns.
:TRIGger:B:PULSEWidth:WHEn ? might return
:TRIGger:B:PULSEWidth:WHEn MOREthan indicating that a trigger is
generated w hen a pulse is detected that is greater than the time specified by
TRIGger:{A|B}:PULSEWidth:LOWLimit.
TRIGger:{A|B}:RUNT:LOGICQUALification
This command specifies whether or not to use logic qualification for a runt trigger.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-859