Commands listed in alphabetical order
Examples
VERTICAL:DESK
EW:FROM:SOURCE CH1
sets channel 1 as the from source.
VERTICAL:DESKEW:FROM:SOUR CE? might r eturn
:VERTICAL:DE
SKEW:FROM:SOURCE CH2
indicating channel 2 is the from
source.
VERTical:DESKEW:STATIC (No Query Form)
This command executes static deskew using the deskew settings.
Group
Vertical
Syntax
VERTical:DESKEW:STATIC EXECute
Arguments
EXECute will execute static deskew using the deskew s
ettings.
Examples
VERTICAL:DESKEW:STATIC EXECUTE will execute static deskew using
the
deskew settings.
VERTical:DESKew:TO:CUSTOMPROPAgation
This command sets or queries a target (TO) delay that can be specied by the user
when the propagation delay of the target (TO) probe used for deskew cannot be
detected automatically.
Group
Vertical
Syntax
VERTical:DESKew:TO:CUSTOM PROPAgation <NR3>
Arguments
<NR3> is a target (TO) delay that can be specied by the user when the propagation
delay of the target (TO) probe used for deskew cannot be detected automatically.
Examples
VERTICAL:DESKEW:TO:CUSTOM PROPAGATION 0.2 5e-9 sets the custom
propagation to 250 ps.
VERTICAL:DESKEW:TO:CUSTOM PROPAGATION? might return
:VERTICAL:DESKEW:TO:CUSTO MPROPAGATION 0.0E+0 indicating the custom
propagation is set to 0.0 ns.
2-888 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals