Commands listed in alphabetical order
Group
Digital
Syntax
CH<x>_D<x>:LABel:NAMe <QString>
Arguments
<QString> is the label.
Examples
CH1_D1:LABel:NAMe "Clock in" sets the name to Clock in.
CH1_D1:LABel:NAMe? might return :CH1_D1: LABEL:NAME "Digital 1 ".
CLEAR (No Query Form)
This command clears acquisitions, measurements, and waveforms.
Group
Miscellaneous
Syntax
CLEAR
Examples
CLEAR clears all acquisitions, measurements, and waveforms.
*CLS (No Query Form)
This command (no query form) clears the following:
Event Queue
Standard Event Status Register
Status Byte Register (except the MAV bit)
If the *CLS command immediately follows an <EOI>, the Output Queue and
MAV bit (Status Byte Register bit 4) are also cleared. MAV indicates that
information is in the output queue. The device clear (DCL) control message
will clear the output queue and thus MAV. *CLS does not clear the output queue
o r M AV.
*CLS can suppress a Service Request that is to be generated by an *OPC. This
will happen if a single sequence acquisition operation i
s still being processed
when the *CLS command is executed.
Group
Status and Error
MSO54, MSO56, MSO58, MSO58LP Programmer 2-211