Commands listed in alphabetical order
TRIGGER:A:BUS
:B1:LIN:DATA:QUALIFIER?
might return
TRIGGER:A:BUS:B1:LIN:DATA :QUALIFIER EQUAL indicating the data
qualifier is set to equal.
TRIGger:{A|
B}:BUS:B<x>:LIN:DATa:SIZe
This command specifies the length o f the data string in bytes to be used for LIN
trigger. The bus number is specified by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:LI N:DATa:SIZe <NR1>
TRIGger:{A|B}:BUS:B<x>:LI N:DATa:SIZe?
Arguments
<NR1> is the size of the data string in bytes.
Examples
TRIGGER:A:BUS:B1:LIN:DATA :SIZE 8 sets the data size to 8 bytes.
TRIGGER:A:BUS:B1:LIN:DATA :SIZE? might return
TRIGGER:A:BUS:B1:LIN:DATA :SIZE 1 indicating the data size is 1 byte.
TRIGger:{A|B}:BUS:B<x>:LIN:DATa:VALue
This command specifies the binary data string to be used for LIN t
rigger condition
if trigger condition is ID or IDANDDATA. The bus number is specified by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:LI N:DATa:VALue <QString>
TRIGger:{A|B}:BUS:B<x>:LI N:DATa:VALue?
Arguments
<QString> is a quoted string that is the LIN trigger data value.
2-812 MSO54, MSO56, MSO58, MSO58LP Programmer