Commands listed in alphabetical order
TRIGger:{A|B
}:TRANsition:DELTatime
This command specifies the delta time (that is Time Limit) used in calculating the
transition value for the transition (that is Rise or Fall Time) trigger.
Group
Trigger
Syntax
TRIGger:{A|B}:TRANsition: DELTatime <NR3>
TRIGger:{A|B}:TRANsition: DELTatime?
Arguments
<NR3> is a floating point number t
hat specifies the delta time, in seconds.
Examples
TRIGGER:A:TRANSITION:DELT ATIME 15E-6 sets the delta time of the
transition trigger to 15 µs.
TRIGGER:A:TRANSITION:DELT ATIME? might return
:TRIGGER:A:TRANSITION:DEL TATIME 2.0000E-09 indicating that the delta
time of the transition trigger is set to 2 ns.
TRIGger:{A|B}:TRANsition:LOGICQUALification
This command specifies whether or not to use logic qualification for a transition
trigger.
Group
Trigger
Syntax
TRIGger:{A|B}:TRANsition: LOGICQUALification {ON|OF F}
TRIGger:{A|B}:TRANsition: LOGICQUALification?
Arguments
ON specifies that the transition trigger type uses logic qualification.
OFF specifies that the transition trigger type does not use logic qualification.
Examples
TRIGger:A:TRANsition:LOGI CQUALification ON specifies using logic
qualification.
TRIGger:A:TRANsition:LOGI CQUALification? might return
:TRIGGER:A:TRANSITION:LOG ICQUALIFICATION OFF indicating logic
qualification is off.
2-868 MSO54, MSO56, MSO58, MSO58LP Programmer