Commands listed in alphabetical order
HORizontal:M
ODe:MANual:CONFIGure
Sets or queries which horizontal c ontrol (scale o r record length) will primarily
change when the sample rate is changed in Manual mode. If the selected control
(scale or rec
ord length) reaches a limit then the unselected control (record length
or scale) may also change.
Group
Horizontal
Syntax
HORizontal:MODe:MANual:CON FIGure {HORIZontalscale|
RECORDLength}
HORizontal:MODe:MANual:CON FIGure?
Arguments
HORIZontalscale will change when sample rate is adjusted.
RECORDLength will change when sample rate is adjusted.
Examples
HORIZONTAL:MODE:MANUAL:CON FIGURE RECORDLength allows the record
length be adjusted when sample rate is changed in Manual m ode.
HORizontal:MODE:RECOrdlength
This command sets or queries the record length.
Group
Horizontal
Syntax
HORizontal:MODE:RECOrdleng th <NR1>
HORizontal:MODE:RECOrdleng th?
Arguments
<NR1> is the record length in samples. Manual mode lets you change the record
length, while the record length is read only for Automatic mode.
Examples
HORIZONTAL:MODE:RECORDLENG TH 1000 sets the record length to
1000 samples.
HORIZONTAL:MODE:RECORDLENG TH? might return
:HORIZONTAL:MODE:RECORDLEN GTH 1000000, indicating that the record
length is set to 1,000,000 samples
MSO54, MSO56, MSO58, MSO58LP Programmer 2-351