9 - 4 BeneVision N1 Patient Monitor Operator’s Manual
9.5 Changing Resp Settings
9.5.1 Setting the Resp Alarm Properties
To set the Resp alarm properties, follow this procedure:
1. Select the Resp numeric area or waveform area to enter the
Resp dialog.
2. Select the
Alarm tab.
3. Enter the password if required. For more information, refer to
21.12.3Selecting Password for User Authentica-
tion.
4. Set alarm properties as desired.
9.5.2 Setting the RR Source
To set RR source, follow this procedure:
1. Select the Resp numeric area or waveform area to enter the
Resp dialog.
2. Select the
Setup tab.
3. Choose
RR Source from the dropdown list.
When you select
Auto, the monitor automatically selects the RR source following order of priority: first CO2, and
then
ECG. When the manually selected RR source is not available, the monitor automatically switches the RR
Source
to Auto.
9.5.3 Choosing the Respiration Lead
To set the respiration lead, follow this procedure:
1. Select the Resp numeric area or waveform area to enter the
Resp dialog.
2. Select the
Setup tab.
3. Set
Resp Lead.
If you cannot get an acceptable impedance Resp waveform or you suspect the veracity of the Resp value after
choosing the Resp lead, you may need to adjust the electrode placement.
9.5.4 Setting the Resp Waveform Size
To set the Resp waveform size, follow this procedure:
1. Select the Resp numeric area or waveform area to enter the
Resp dialog.
2. Select the
Setup tab.
3. Set
Gain.
9.5.5 Setting the Resp Waveform Speed
To set the Resp waveform speed, follow this procedure:
1. Select the Resp numeric area or waveform area to enter the
Resp dialog.
2. Select the
Setup tab.
3. Set
Speed.