21 - 4 BeneVision N1 Patient Monitor Operator’s Manual
4. Select Increase Volume Delay to set the delay time of alarm volume escalation.
NOTE
• The alarm volume escalation function is not applied to the latched alarms.
21.3.2 Setting Alarm Pausing and Resetting Properties
21.3.3 Defining the Pause Function
You can either pause alarms or pause alarm tones. This depends on the pause setting. To set the pause function,
follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab → Pause/Reset tab.
3. Set
Pause to Alarm Pause or Audio Pause. The default setting is Audio Pause.
21.3.3.1 Setting the Alarm Pause Time/Alarm Tone Pause Time
The alarm pause time or alarm tone pause time can be set to 1 min, 2 min, 3 min, or Permanent. The default
audio pause time is two minutes.
To set the alarm tone pause time, follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab → Pause/Reset tab.
3. Set
Pause Time.
21.3.3.2 Setting the Priority of Audio Paused Alarms
To select alarm of what priority can be paused or alarm sound of what priority can be paused, follow this
procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab → Pause/Reset tab.
3. Set
Pause Priority.
â—† All: pressing the Alarm Pause quick key pauses all alarms.
â—† Med & Low: pressing the Alarm Pause quick key pauses alarms of medium and low priority. The high
priority alarms will not be paused.
â—† Disable: the Alarm Pause quick key is disabled.
21.3.3.3 Disabling the Options of Prolonging Alarm Pause Time/Alarm Tone Pause Time
To disable specific options for the clinician when prolonging alarm pause time or the alarm tone pause time,
follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab → Pause/Reset tab.
3. Switch off
Pause 5 min, Pause 10 min, or Pause 15 min.

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals