B - 4 BeneVision N1 Patient Monitor Operator’s Manual
B.2 Radio Regulatory Compliance
RF parameters for the Wi-Fi module
This device complies with Part 15 of the FCC Rules and RSS-210 of Industry Canada. Operation is subject to the
following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
Any changes or modifications to this equipment not expressly approved by Mindray may cause harmful radio
frequency interference and void your authority to operate this equipment.
The maximum antenna gain permitted complies with the e.i.r.p. limits as stated in RSS-210.
The maximum antenna gain permitted complies with the e.i.r.p. limits specified for point-to-point operation, as
stated in RSS-210.
Rated Maximum
Output power of
Transmitter Watts
(W)
Separation Distance According to Frequency of Transmitter (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.7 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.20 2.3
10 3.8 3.80 7.3
100 12 12.00 23
For transmitters at a maximum output power not listed above, the recommended separation distanced in meters (m) can
be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in Watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Recommended Separation Distances between Portable and Mobile RF, Communications Equipment and This
Equipment
d
3.5
V
-------
P=
d
3.5
E
-------
P=
d
7
E
---
P=
Item IEEE 802.11b/g/n (2.4G) IEEE 802.11a/n (5G)
Operating Frequency ETSI: 2.4 GHz to 2.483 GHz
FCC: 2.4 GHz to 2.483 GHz
MIC: 2.4 GHz to 2.495GHz
KC: 2.4 GHz to 2.483 GHz
ETSI: 5.15 GHz to 5.35 GHz, 5.47 GHz to 5.725 GHz
FCC:5.15GHz to 5.35GHz, 5.47~5.725GHz to
5.725GHz to 5.825GHz
MIC: 5.15GHz to 5.35GH, 5.47 to 5.725GHz
KC:5.15GHz to 5.35GHz, 5.47 to 5.725GHz, 5.725GHz
to 5.825GHz
Modulation Mode DSSS and OFDM OFDM
Output power <30dBm (Peak Power)
<20dBm (Average Power)

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals