BeneVision N1 Patient Monitor Operator’s Manual 6 - 7
6.6.5 Setting the Zero RR Delay Time
To set the zero respiration rate delay time, follow this procedure:
1. Select the Main Menu quick key → from the Alarm column select Setup.
2. Select Zero RR to set the zero respiration rate delay time.
6.6.6 Restoring the Default Alarm Settings
To reset all alarm settings to the defaults, follow this procedure:
1. Select the Main Menu quick key → from the Alarm column select Limits.
2. On the Limits page, select Defaults at the bottom.
6.6.7 Setting the Length of Printed Waveforms
You can define the length of printed waveforms when an alarm is triggered. To do so, follow this procedure:
1. Select the Main Menu quick key → from the Alarm column select Setup.
2. Set Printing Duration On Alarm.
6.7 Pausing Alarms/Pausing Alarm Tones
You can either pause alarms or pause alarm tones. This depends on the pause setting. To set the pause function,
refer to 21.3.3Defining the Pause Function.
6.7.1 Pausing Alarms
If the Pause function is configured to alarm pause, pressing the Alarm Pause quick key can temporarily disable
alarm indicators. For the configuration of the pause function, see section 21.3.3Defining the Pause Function.
When alarms are paused, the following rules are followed:
â– No physiological alarm will be presented.
CO
2
EtCO
2
(mmHg)
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
Same as the
measurement range
33 to 35: 29 33 to 35: 29 33 to 35: 41 33 to 35: 41 Same as the
measurement range
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
+ 6)
36 to 45:
(etCO
2
+ 6)
Same as the
measurement range
46 to 48: 39 46 to 48: 39 46 to 48: 51 46 to 48: 51 Same as the
measurement range
>48: remains
the same
>48: remains
the same
>48: remains
the same
>48: remains
the same
Same as the
measurement range
FiCO
2
None None Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
awRR
(rpm)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
(awRR+25) or
85 rpm
(whichever is
smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
Module Parameter Lower Limit Upper Limit Auto Limit Range
Adult/
pediatric
Neonate Adult/
pediatric
Neonate