BeneVision N1 Patient Monitor Operator’s Manual 21 - 7
2. Select the Alarm tab select the Remote View tab.
3. Set
Alarm Sound to ISO, Mode 1 or Mode 2.
21.3.5.6 Switching Off the Remote Device Disconnection Alarm
The monitor can provide an alarm if remote devices, for example, a bedside monitor or a telemetry, are
disconnected. By default, the function is enabled. To disable the alarm, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Remote View tab.
Switch off
Remote Disconnected Alarm.
21.3.6 Changing Alarm Priority Settings
21.3.6.1 Setting the Priority of the ECG Lead Off Alarm
To set the priority of the ECG lead off alarm, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab select the Other tab.
3. Set the priority of
ECG Lead Off.
21.3.6.2 Setting the Priority of the SpO
2
Sensor Off Alarm
To set the priority of the SpO
2
sensor off alarm, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab select the Other tab.
3. Set the priority of
SpO2 Sensor Off.
21.3.6.3 Setting the Priority of the IBP No Sensor Alarm
To set the priority of the IBP no sensor alarm, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab select the Other tab.
3. Set the priority of
IBP No Sensor.
21.3.6.4 Setting the Switch of the CMS and eGateway Disconnection Alarm
You can choose whether to issue an alarm when the monitor is not connected or disconnected from the CMS/
eGateway. To do so, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Other tab.
3. In the
Other block, switch on or off CMS/eGW Disconnected Alarm. If CMS/eGW Disconnected Alarm is
switched off, the “Offline” alarm is not presented when the monitor is not connected or disconnected from
the CMS/eGateway.
21.3.6.5 Setting the Priority of the CMS and eGateway Disconnection Alarm
To set the priority of the CMS and eGateway disconnection alarm, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals