BeneVision N1 Patient Monitor Operator’s Manual 14 - 3
If the sidestream CO
2
module is configured with the oxygen sensor, O
2
waveform and parameters can be
displayed in an O
2
numeric and waveform area as follows:
14.6 Measuring CO
2
Using the Sidestream/Microstream CO
2
Module
14.6.1 Preparing to Measure CO
2
Using the Sidestream CO
2
Module
To prepare the sidestream CO
2
measurement, follow this procedure:
1. Select the appropriate gas sample line and watertrap according to the patient category.
2. Connect one end of the gas sample line to the watertrap or to the CO
2
adapter.
â—† If using the sidestream CO
2
module, connect the watertrap to the CO
2
module, and connect the gas
sample line to the watertrap.
â—† If using the N1 monitor for CO
2
measurement, connect one end of the gas sample line to the CO
2
adapter. Refer to the
CO2 Adapter User Manual (PN: H-046-009994-00) for the connection of the CO
2
adapter and the gas sampling line.
(1) CO
2
waveform (2) End tidal CO
2
value (EtCO
2
)
(3) Airway respiration rate (awRR) (4)Fraction of inspired CO
2
(FiCO
2
)
(1) O
2
waveform (2) FiO
2
measurement (3) EtO
2
measurement
(1) Watertrap receptacle (2) DRYLINE II watertrap
(3)
Gas sample line