12 - 4 BeneVision N1 Patient Monitor Operator’s Manual
arm or the thigh. The inflatable part of the cuff should be long enough to encircle to overlap at least
50% to 80% of the limb.
c Apply the cuff to the patient’s upper arm or leg and make sure the Φ marking on the cuff matches the
artery location. The cuff should fit snugly, but with enough room for two fingers to be placed
between the cuff and the patient
’s arm (on adults), and loosely on neonates with little or no air
present within the cuff. Otherwise it may cause discoloration and ischemia of the extremities. Make
sure that the cuff index line falls within the range markings on the cuff. If it does not, use a cuff that fits
better.
d Middle of the cuff should be at the level of the right atrium of the heart. If it is not, use the
measurement correction formula to correct the measurement. For more information, see
12.8.10Correcting the NIBP Measurements.
4. Connect the cuff to the air tubing. Avoid compression or restriction of pressure tubes. Air must pass
unrestricted through the tubing.
CAUTION
• Do not touch or apply external pressure against the cuff and air tubing during NIBP measurement.
This may cause inaccurate blood pressure values.
• Use care when placing the cuff on an extremity used for monitoring other patient parameters.
12.7 Starting and Stopping NIBP Measurements
Start and stop NIBP measurement by selecting the NIBP quick keys or from the NIBP dialog.
12.8 Changing NIBP Settings
12.8.1 Setting the NIBP Alarm Properties
To set the NIBP alarm properties, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP dialog.
2. Select the
Alarm tab.
Task By Quick Key From NIBP dialog
Start a manual measurement NIBP Start/Stop quick key Start NIBP button
Start auto NIBP series
NIBP Start/Stop quick key Setup tab → set Interval → Start NIBP
button
NIBP Measure quick key →select
interval
Start NIBP sequence
measurement
NIBP Measure quick key →
Sequence
Sequence tab → set NIBP sequence
→Start NIBP button
Start STAT measurement
NIBP Measure quick key → STAT STAT button
Stop the current NIBP
measurements
NIBP Start/Stop quick key Stop NIBP button
End auto NIBP series or NIBP
Sequence
/Stop All button
Stop STAT measurement and end
series
NIBP Start/Stop quick key Stop NIBP or Stop All button

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals