BeneVision N1 Patient Monitor Operator’s Manual C - 11
C.7 CO
2
Default Settings
C.7.1 General Settings
Scale ICP/RAP/LAP/UVP
venous pressure
0-20 mmHg
Art/Ao/BAP/FAP/LV/P1/
P2 arterial pressure
0-160 mmHg
UAP/P3/P4 venous
pressure
0-80
mmHg
CVP/PA CVP: 0-30cmH
2
O
PA: 0-30 mmHg
Display Format Sys/Dia (Mean)
Display Alarm Limits Off
Use PA-D as PAWP (only available for
independent external display)
Off
PPV Measure Off
PPV Source Auto
PAWP Reference Waveform 1 II
Reference Waveform 2 Resp
Speed 12.5 mm/sec
PA Scale 0-30 mmHg
Overlapping
Waveform Setup
Left Scale 0-160 mmHg
Right Scale 0-20 mmHg
CVP Scale 0-20 cmH
2
O
ICP Scale 0-30 mmHg
PA Scale 0-30 mmHg
Speed 25 mm/sec
Gridlines Off
Item Default Setting
Item Default Setting
EtCO
2
Alarm switch On
High limit Adult and pediatric: 50 mmHg
Neonate: 45 mmHg
Low limit Adult and pediatric: 25mmHg
Neonate: 30mmHg
Priority Med
Alarm Outputs Off
FiCO
2
Alarm switch On
High limit 4 mmHg
Priority Med
Alarm Outputs Off

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals