BeneVision N1 Patient Monitor Operator’s Manual D - 7
D.2.9 Power Supply
Technical Alarm Messages
XX represents 2.5, 3.3,5, or 12.
D.2.10 Printer Technical Alarm Messages
CO
2
Airway Error Low C 1. Check if the sample line is kinked or occluded.
2. Replace the sample line.
3. Reinsert the module.
4. If the alarm persists, contact your service
personnel.
CO
2
Adapter Error Low A Check, clean or replace the airway adapter.
Perform a zero calibration.
CO
2
No Sensor Low A Make sure that the CO
2
transducer is connected.
CO
2
: Change Watertrap Low C Replace the watertrap.
CO
2
Watertrap and
Patient Mismatch
Low C Check the patient category and use a correct
watertrap.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Low Battery Med C Connect the monitor to an AC power source and
allow the batteries to charge.
Critically Low Battery High C Connect the monitor to an AC power source and
allow the batteries to charge.
Battery Service Required Low B The battery has reached its lifetime. Contact your
service personnel.
Power Board Comm Error High C Shutdown and restart the monitor. If the alarm
persists, contact your service personnel.
Battery Error High C The battery may have failed. Contact your service
personnel.
Battery Charging Error High C The charging circuit fails or the battery fails.
Contact your service personnel.
Battery Temperature Too
High
High C Stop using the monitor after this alarm appears,
and contact your service personnel.
Battery Off High C Shutdown and restart the monitor. If the alarm
persists, contact your service personnel.
RT Clock Need Reset High C Contact your service personnel.
RT Clock Not Exist High C Contact your service personnel.
XX V Too High High C There is a problem with the system power supply.
Shutdown and restart the monitor.
XX V Too Low High C
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Printer Buffer Full Prompt / The printer buffer is full. Wait untill the printer
finishes the printing task.
Printing Failed Prompt / The printer ran out of paper or cannot be found
on the network. Check the printer.

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals