BeneVision N1 Patient Monitor Operator’s Manual D - 1
D Alarm Messages
This chapter lists only the most important physiological and technical alarm messages. Some messages
appearing on your monitor may not be included. In the “Cause and solution” column, corresponding solutions
are given instructing you to troubleshoot problems. If the problem persists, contact your service personnel.
D.1 Physiological Alarm Messages
D.1.1 General Physiological Alarm Messages
Note: XX represents a measurement or parameter label, such as HR, NIBP, PVCs, RR, SpO
2
, PR, etc.
D.1.2 Arrhythmia Alarm Messages
Alarm messages Default priority Cause and solution
XX High Med XX value has risen above the high alarm limit or fallen below the low
alarm limit. Check the patient’s condition and check if the patient
category and alarm limit settings are correct.
XX Low Med
Alarm message Default priority
Asystole High
V-Fib/V-Tach High
V-Tach High
Vent Brady High
Extreme Tachy High
Extreme Brady High
PVCs/min Med
Pauses/min Med
R on T Med
Bigeminy Med
Trigeminy Med
Tachy Med
Brady Med
Multiform PVC Med
Vent Rhythm Med
Nonsus V-Tach Med
Run PVCs Low
Pause Low
Couplet Prompt
PVC Prompt
Irr Rhythm Prompt