C - 8 BeneVision N1 Patient Monitor Operator’s Manual
C.6 IBP Default Settings
Initial Pressure Adult: 160 mmHg
Pediatric: 140 mmHg
Neonate: 90 mmHg
Interval OR: 5 min
Neonatology: 30 min
Other departments: 15 min
Start Mode Clock
NIBP End Tone Off
Venipuncture Pressure Auto
Display Format Sys/Dia (Mean)
Display Alarm Limits Off
Display PR Off
Item Default Setting
Item Default Setting
IBP-S Alarm switch On
High limit • Art/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 160 mmHg
Pediatric: 120 mmHg
Neonate: 90 mmHg
•PA
Adult: 35 mmHg
Pediatric and neonate: 60 mmHg
Low limit • Art/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 55 mmHg
•PA
Adult: 10 mmHg
Pediatric and neonate: 24 mmHg
Priority Med
Alarm Outputs Off