BeneVision N1 Patient Monitor Operator’s Manual D - 5
D.2.6 NIBP Technical Alarm Messages
D.2.7 IBP Technical Alarm Messages
Note: XX represents an IBP label, for example PA, CVP, FAP, P1, etc.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
NIBP Cuff Loose Low A There is a leak in the cuff or air tubing. Use a cuff
of correct type based on the patient size. Apply
the cuff and connect the air tubing as instructed
in the manual.
NIBP Cuff or Airway Leak Low A Check the NIBP cuff and pump for leakages.
NIBP Airway Error Low A The air tubing may be occluded. Check the air
tubing for an occlusion or kinking. If the alarm
persists, contact your service personnel.
NIBP Weak Signal Low A The patient’s pulse is weak or the cuff is loose.
Check the patient’s condition and replace the cuff
application site.
NIBP Overrange Low A The measured NIBP value exceeds the module
measurement range. Check the patient’s
condition.
NIBP Excessive Motion Low A Check the patient’s condition and reduce patient
motion.
NIBP Cuff Overpressure Low A The NIBP airway may be occluded. Check the
airway and measure again. If the alarm persists,
contact your service personnel.
NIBP Timeout Low A The measurement time exceeds 120 seconds in
the adult or pediatric mode, or exceeds 90
seconds in the neonatal mode, and the BP value
cannot be obtained. Check the patient’s
condition and NIBP connections, or replace the
cuff and measure again.
NIBP Cuff and Patient
Mismatch
Low A The cuff type mismatches the patient category.
Verify the patient category or replace the cuff if
necessary. If patient category is correct, check
that the tubing is not bent and the airway is not
occluded.
NIBP Airway Leak Low A Airway leakage is found during the NIBP leakage
test. Check the NIBP cuff and pump for leakages.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
XX Sensor Error Med C The IBP sensor fails. Replace the sensor.
XX No Sensor Med A The IBP patient cable and/or corresponding IBP
sensor is not connected or detached. Check the
cable and sensor connection.
XX No Pulse Low A The catheter may be occluded. Please flush the
catheter.
XX Disconnected High C The tubing is disconnected from the patient, or
the three-way valve is open to the air. Check the
connection of the tubing, or check the valve is
open to the patient. If the alarm persists, contact
your service personnel.

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals