C - 4 BeneVision N1 Patient Monitor Operator’s Manual
C.1.4 QT Default Settings
C.1.5 Glasgow 12-lead ECG Algorithm Default Settings
C.2 Respiration Default Settings
Item Default Setting
QTc Alarm switch Off
High limit Adult: 500
Pediatric: 480
Neonate: 460
Priority Med
Alarm Outputs Off
ΔQTc Alarm switch Off
High limit 60
Priority Med
Alarm Outputs Off
QT Analysis Off
QT Lead All
Item Default Setting
Filter 35 Hz
Baseline Drift Removal On
Tachy 100
Brady 50
Waveform Layout Standard
Median Complex Off
Measurements On
Interpretation On
Interpretation Summary On
Item Default Setting
RR Alarm switch On
High limit Adult: 30
Pediatric: 30
Neonate: 100
Low limit Adult: 8
Pediatric: 8
Neonate: 30
Priority Med
Alarm Outputs Off