BeneVision N1 Patient Monitor Operator’s Manual 20 - 3
4. Select Export Configuration.
5. Select the configurations and
User Maintenance Settings to export.
6. Select
Export.
20.7.2 Importing a Configuration
To import the configuration from the USB drive to the monitor, follow this procedure:
1. Connect the N1 to the Dock.
2. Connect the USB drive to the Dock’s USB port.
3. Select the
Main Menu quick key → from the Configuration column select Manage → input the required
password
→ Select .
4. Select
Import Configuration.
5. Select the configurations and
User Maintenance Settings to import.
6. Select
Import.
20.7.3 Loading a Configuration
You may make changes to some settings during operation. However, these changes or the pre-selected
configuration may not be appropriate for the newly admitted patient. Therefore, the monitor allows you to load
a desired configuration to ensure that all the settings are appropriate for your patient.
To load a configuration, follow this procedure:
1. Select the
Main Menu quick key → from the Configuration column select Load.
2. Select the desired configuration.
â—† Select the configuration on this monitor in the Local page.
â—† Select the configuration on the USB drive in the USB Drive page.
3. Select
Load.
• The monitor may configure some settings by default when you load a configuration of different
software version with the current configuration.
20.8 Printing Configurations
To print both factory configurations and user configurations, follow this procedure:
1. Select the
Main Menu quick key →from the Configuration column select Manage → input the required
password
→ select .
2. Select
Print Configuration.
3. Select desired configurations.
4. Select
Print.
20.9 Modifying Configuration Password
To modify the configuration password, follow this procedure:
1. Select the
Main Menu quick key → from the Configuration column select Manage → input the required
password
→ Select .
2. Select
Modify Password.
3. Respectively input the old password and new password.
4. Select
OK.