A - 16 BeneVision N1 Patient Monitor Operator’s Manual
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured
with this device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2:2013)
in terms of mean error and standard deviation by comparing with intra-arterial or auscultatory measurements
(depending on the configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff
sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the
American National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and
standard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical
patient population.
A.14.7 IBP Specifications
Static pressure measurement range 0 mmHg to 300 mmHg
Static pressure measurement accuracy ±3 mmHg
Alarm limit Range (mmHg)
Step (mmHg)
NIBP-S High Adult: (low limit + 5) to 285
Pediatric: (low limit + 5) to 235
Neonate: (low limit + 5) to 135
NIBP ≤ 50: 1
NIBP > 50: 5
NIBP-S Low 26 to (high limit - 5)
NIBP-M High Adult: (low limit + 5) to 255
Pediatric: (low limit + 5) to 210
Neonate: (low limit + 5) to 120
NIBP-M Low 16 to (high limit - 5)
NIBP-D High Adult: (low limit + 5) to 245
Pediatric: (low limit + 5) to 195
Neonate: (low limit + 5) to 110
NIBP-D Low 11 to (high limit - 5)
NIBP-S Extreme High Adult: (NIBP-S high limit + 5) to 290
Pediatric: (NIBP-S high limit +5) to 240
Neonate: (NIBP-S high limit +5) to 140
NIBP ≤ 50: 1
NIBP > 50: 5
NIBP-S Extreme Low 25 to (NIBP-S low limit - 5)
NIBP-M Extreme High Adult: (NIBP-M high limit + 5) to 260
Pediatric: (NIBP-M high limit + 5) to 215
Neonate: (NIBP-M high limit + 5) to 125
NIBP-M Extreme Low 15 to (NIBP-M low limit - 5)
NIBP-D Extreme High Adult: (NIBP-D high limit + 5) to 250
Pediatric: (NIBP-D high limit + 5) to 200
Neonate: (NIBP-D high limit + 5) to 115
NIBP-D Extreme Low 10 to (NIBP-D low limit - 5)
Standard Meets the requirements of IEC 60601-2-34: 2011.
Technique Direct invasive measurement
IBP
Measurement range -50 to 300 mmHg
Resolution 1 mmHg
Accuracy ±2% or ±1 mmHg, whichever is greater (excluding sensor error)
Refresh rate ≤1s