23.3.3 Disinfecting the Equipment and Mounting Kits
Disinfect the main unit, parameter modules, Modular Rack, Dock, folding hook, monitor handle, and bedrail
hook as required in your hospital’s servicing schedule using the disinfecting agents listed in the table above.
Cleaning equipment and mounting kits before disinfecting is recommended.
CAUTION
• Never use EtO or formaldehyde for disinfection.
23.4 Cleaning and Disinfecting the Accessories
For the NIBP air hose, Masimo SpO
2
cable, and Nellcor SpO
2
cable, you should clean and disinfect them using the
cleaners and disinfectants and methods listed in this section. For other accessories, you should consult the
instructions delivered with the accessories.
CAUTION
• Fluids entering the NIBP air hose can damage the equipment. When cleaning and/or disinfecting the
NIBP air hose, prevent liquid from entering the hose.
• Periodically inspect the NIBP air hose and connector for signs of wear or deterioration. Replace the
NIBP air hose if you detect a leak. Dispose of damaged NIBP air hose according to local laws for
disposal of hospital waste.
• Never immerse or soak the accessories in any liquid.
• Never clean or disinfect the metalic connectors at either end of the accessories.
• Use only Mindray approved cleaners and disinfectants and methods listed in this section to clean or
disinfect the accessories. Warranty does not cover damage caused by unapproved substances or
methods.
• To avoid long term damage, the accessories should be disinfected only when necessary as
determined by your hospital’s policy.
23.4.1 Approved Accessories Cleaning and Disinfecting Agents
The following table lists approved NIBP air hose cleaning and disinfecting agents:
:
Product Name Product Type Active Ingredients
Isopropanol Liquid Isopropanol 70%
1-Propanol 1-Propanol 50%
CIDEX® OPA Ortho-Phthalaldehyde 0.55%
Metrex CaviCide1â„¢ Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium
chloride 0.28%,
Isopropanol 17.2%
Virex® TB n-Alkyl dimethyl benzyl ammonium chlorides 0.105%,
n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.105%
Rely+On™ Virkon® High Level
Surface Disinfectant
Powder Used as 1% solution
Biocidal active:
Pentapotassium bis (peroxymonosulphate) bis (sulphate)(500g/
kg),
Contains dipotassium peroxodisulphate.

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals