10 - 8 BeneVision N1 Patient Monitor Operator’s Manual
10.9 Masimo Information
Masimo Patents
This posting serves as notice under 35 U.S.C.§287(a) for Masimo patents: http://www.masimo.com/patents.htm.
No Implied License
Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of
one or more of the patents relating to this device.
10.10 Nellcor Information
Nellcor Patents
This posting serves as notice under 35 U.S.C.§287(a) for Covidien patents: http://www.covidien.com/patents.
No Implied License
Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized replacement parts which would, alone, or in combination with this device, fall within the scope of
one or more of the patents relating to this device.
Dashes “- -” display in place of numerics 1. Check that the cable connections of SpO
2
sensor and the
extension cable are tight. Replace the SpO
2
sensor or the
extension cable if needed.
2. Reconnect the SpO
2
sensor if the alarm SpO2 Sensor Off
appears.
3. Check the PI value. If the PI value is too low, adjust the SpO
2
sensor, or apply the sensor to a site with better perfusion.
4. Move the sensor to a place with weaker ambient light, or cover
the sensor to minimize the ambient light if the alarm SpO2 Too
Much Light
appears.
Low amplitude SpO
2
signal 1. The SpO
2
sensor and NIBP cuff are placed on the same limb.
Change the monitoring site if necessary.
2. Check the PI value. If the PI value is too low. Adjust the SpO
2
sensor, or apply the sensor to a site with better perfusion.
SpO
2
value is inaccurate 1. Check the patient’s vital signs.
2. Check for conditions that may cause inaccurate SpO
2
readings.
For more information, see
10.3SpO
2 Measurement Limitations
.
3. Check the monitor and the SpO
2
sensor for proper functioning.
Problem Solution

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals