6 - 2 BeneVision N1 Patient Monitor Operator’s Manual
6.3.2 Alarm Priorities
By severity, the alarms are classified into the following priority levels:
â–  High priority alarms: indicates a life threatening situation or a severe device malfunction. High priority
alarms require an immediate response.
â–  Medium priority alarms: indicates abnormal vital signs or a device malfunction. Medium priority alarms
require a timely response.
â–  Low priority alarms: indicate a discomfort condition, a device malfunction, or an improper operation. Low
priority alarms require you to be aware of this condition.
â–  Messages: provides additional information on the patient or the equipment.
6.3.3 Alarm Indicators
When an alarm occurs, the monitor indicates it through visual or audible alarm indications. For more
information, see the following table.
NOTE
• When multiple alarms of different priority levels occur simultaneously, the monitor selects the
highest priority alarm to light the alarm lamp and issue the alarm tone.
• When multiple alarms of different priority levels occur simultaneously and should be displayed in
the same area, the monitor only displays the messages from the highest priority alarm.
• When multiple alarms of the same priority levels occur simultaneously and should be displayed in
the same area, all the alarm messages are displayed cyclically.
• Some physiological alarms, such as Asystole, are exclusive. They have identical alarm tones and
alarm lights with normal high level physiological alarms, but their alarm messages are displayed
Alarm Indicator High Priority
Alarm
Medium
Priority Alarm
Low Priority
Alarm
Message Comments
Alarm lamp Red
Flashing
frequency: 1.4 -
2.8 Hz
Duty cycle: 20 -
60% on
Yellow
Flashing
frequency: 0.4 -
0.8 Hz
Duty cycle: 20 -
60% on
Cyan
No flashing
Duty cycle:
100% on
None None
Audible
tone
pattern
ISO Repeat pattern
of triple +
double + triple
+ double beeps
Repeat pattern
of triple beeps
single beep None None
Mode 1 Repeat pattern
of high-pitched
single beep
Repeat pattern
of double beeps
Low-pitched
single beep
None
Mode 2 Repeat pattern
of high-pitched
triple beeps
Repeat pattern
of double beeps
Low-pitched
single beep
None
Alarm message White text
inside a red box
Black text inside
a yellow box
Black text inside
a cyan box
White
text
Alarm messages are
displayed in the alarm
information area at the
top of the screen. You
can select the alarm
messages to show the
alarm list.
Alarm priority
indicator
*** ** * None The indicator appears in
front of the
corresponding alarm
message.
Parameter value White text
inside a flashing
red box
Black text inside
a flashing
yellow box
Black text
inside a flashing
cyan box
None None

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals