A - 14 BeneVision N1 Patient Monitor Operator’s Manual
A.14.4 PR Specifications
PR from Masimo SpO
2
PR from Nellcor SpO
2
PR from NIBP
PR from IBP
Alarm limit Range (bpm) Step (bpm)
PR High PR≤40: (low limit +2) to 40
PR>40: (low limit +5) to 295
PR≤40: 1
PR>40: 5
PR Low PR≤40: 16 to (high limit - 2)
PR>40: 40 to (high limit - 5)
Measurement range 25 to 240 bpm
Resolution 1 bpm
Response time ≤20 s (with normal perfusion, no disturbance, and a PR value transition from
25 to 220 bpm)
Accuracy ±3 bpm (without motion)
±5 bpm (with motion)
Refresh rate ≤1 s
Measurement range 20 to 300 bpm
Resolution 1 bpm
Response time ≤30 s (with normal perfusion, no disturbance, and a PR value transition from
25 to 250 bpm)
Accuracy 20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refresh rate ≤1 s
Measurement range 30 to 300 bpm
Resolution 1 bpm
Accuracy ±3 bpm or ±3%, whichever is greater
Measurement range 25 to 350 bpm
Resolution 1 bpm
Accuracy ±1 bpm or ±1%, whichever is greater