BeneVision N1 Patient Monitor Operator’s Manual 21 - 5
21.3.3.4 Setting Alarm Light Status on Alarm Reset
When the alarm system is reset, the monitor presents the alarm light by default , but you can switch off the alarm
light. To change the alarm light setting, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Pause/Reset tab.
3. Set
Alarm Light.
On When Reset: when the alarm system is reset, the alarm tones of the current alarms are switched
off, but the alarm lamp remains flashing.
Off When Reset: when the alarm system is reset, both the alarm tone and alarm lamp of the current
alarms are switched off.
21.3.3.5 Setting the Reminder Tones
When the alarm volume is set to zero, or the alarm is reset or switched off, the monitor issues a periodical
reminder tone. The reminder tone is enabled by default.
To set the reminder tone, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Pause/Reset tab.
3. Set
Alarm Reset Reminder and Alarm Off Reminder.
On: the monitor issues reminder tones at a designated interval.
Re-alarm: if the alarm condition persists the acknowledged alarms marked with “√” will be re-
generated after the designated reminder tone interval.
Off: the monitor does not issue reminder tones at a designated interval. The acknowledged alarms
marked with “√ will be silenced.
4. Set Reminder Interval to 10 min, 5 min, 3 min, 2 min, or 1 min.
21.3.4 Latching Physiological Alarms
To latch physiological alarms, follow this procedure:
1. Select the
Main Menu quick key from the System column select Maintenance input the required
password
select .
2. Select the
Alarm tab Latching tab.
3. Select how you want to latch the alarms.
Alarm latching rules are as follows:
You can separately latch visible alarm signals and audible alarm signals.
Latching audible alarm signals simultaneously latches the corresponding visible signals.
Selecting alarms of lower priority simultaneously latches higher priority alarms.
Switching V-Tach Latching Lock off allows you to switch on or off V-Tach Latching.
When V-Tach Latching Lock is switched on, V-Tach latching follows the setting of Lethal alarm latching and
you cannot control V-Tach Latching.
21.3.5 Changing Remote Alarm Settings
21.3.5.1 Resetting Alarms for Remote Devices
You can reset the alarms occurring on the remote devices that are viewed on the Remote View screen of the
monitor. To enable this function, follow this procedure:

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals