15 - 6 BeneVision N1 Patient Monitor Operator’s Manual
15.2.7.4 Viewing Event Details
To view waveforms and parameter values at the event time, follow this procedure:
1. Enter the
Events page.
2. Select
Detail.
To display beat labels on the first ECG waveform, switch on
Beat Anno. The white beat labels indicate heart beat
classification and may explain suspected, missed, or false arrhythmia calls. Heart beats are classified as follows:
â– N = Normal
â– V = Ventricular ectopic
â– S = Supraventricular premature
â– P = Paced
â– L = Learning
â– ? = Insufficient information to classify beat
â– I = Inoperative (for example, Lead Off)
â– M = Missed beat
If you switch on
Beat Anno on the Events page, beat labels will also be displayed on the Full Disclosure page,
and vise versa. Beat labels can be printed out.
15.2.7.5 Printing Event Reports
To print event reports via a printer, follow this procedure:
1. Enter the events review page.
2. Select to enter the
Print Setup dialog.
3. Select the desired options.
â—† Print All Event List: print the entire event list.
â—† Print List of Selected Events: print the list of selected events.
â—† Print Detail of Selected Events: print the details of selected events.
â—† Print Displayed Event Detail: print the waveforms and parameters of the currently displayed event.
4. Select
Print.
15.2.8 Full Disclosure Review Page
You can review up to 48-hours of waveform data on the full disclosure review page. To enter the full disclosure
review page, select the
Main Menu quick key → from the Review column select Full Disclosure.
15.2.8.1 Selecting Waveforms
You can select the waveforms to be displayed and stored. Up to 48 hours of waveforms for all parameters can be
stored. When closing the
Select Waveform dialog, a prompt will inform you of the duration of waveforms that
can be saved if the amount is less than 48 hours. To save and display the desired waveforms, follow this
procedure:
1. Enter the full disclosure review page.
2. Select
and select Setup to enter the Select Waveform page.
3. Select the
Storage tab and set the desired waveforms to be stored in the monitor. Select the Display
(Maximum: 3)
tab and set the desired waveforms to be displayed on the Full Disclosure page.
• Storing more waveforms may shorten the storage time to less than 48 hours. Please adjust the
waveforms stored with this consideration.
In case of alarms, the background of the compressed waveform is highlighted with a colored block during the
alarm time:
â– Red: high alarm priority