15 - 4 BeneVision N1 Patient Monitor Operator’s Manual
NOTE
• Changing between intervals does not typically change the position of the slider or cursor except to
round it to the nearest interval, however changing intervals to be 30 seconds or less will move any
cursor position more than 4 hours ago to the oldest data less than 4 hours ago.
15.2.5.4 Printing a Tabular Trends Report
To print a tabular trends report, follow this procedure:
1. Enter the tabular trends review page.
2. Select to enter the
Print Setup dialog.
3. Set the tabular trends report as described in
18.6.3Configuring Tabular Trends Reports.
4. Select
Print or Print (All).
â—† Print: the trend data before the above set Time, for the above set Period, and at the above set
Interval will be printed.
â—† Print (All): all the stored trend data will be printed at the review interval.
15.2.6 Graphics Trends Review Page
The graphic trends review page displays trend data in a visual format. To enter the graphic trends review page,
select the
Main Menu quick key → from the Review column select Graphic Trends.
15.2.6.1 Changing the Graphic Trend Group
To change the graphic trend group, follow this procedure:
1. Enter the Graphic trends review page.
2. Select and set
Trend Group.
15.2.6.2 Editing the Graphic Trend Group
You can edit the graphic trend group by following this procedure:
1. Enter the Graphic trends review page.
2. Select and select Group Setup.
3. Refer to section
15.2.5.1Editing the Tabular Trend Group for detail on how to edit the group.
15.2.6.3 Changing the Resolution of Trend Data
To change the length of trend data displayed on the current screen, follow this procedure:
1. Enter the graphic trends review page.
2. Select
Zoom.
â—† 8 min: the screen displays eight minutes of trend data. You can view the most recent one hour of
data.
â—† 30 min, 1 hr, 2 hrs, 4 hrs: the screen displays 30 minutes, one hour, two hours, or four hours of trend
data. You can view the most recent four hours of data.
â—† 8 hrs, 12 hrs, 24 hrs, 48 hrs: the screen displays 8 hours, 12 hours, 24 hours, or 48 hours of trend data.
You can view the most recent 120 hours of data.
15.2.6.4 Changing the Number of Waveforms
To change the number of waveforms displayed on the trend review page, follow this procedure:
1. Enter the graphic trends review page.
2. Select
and set Trends.

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals