A - 2 BeneVision N1 Patient Monitor Operator’s Manual
A.3 Environmental Specifications
• The monitor may not meet the performance specifications if stored or used outside the specified
temperature and humidity ranges. If the performance of the equipment is degraded due to aging or
environmental conditions, contact your service personnel.
• When the monitor and related products have differing environmental specifications, the effective
range for the combined products is that range which is common to the specifications for all
products.
• The monitor cannot be transported in the temperature lower than -30ºC.
• The environmental specification of unspecified parameter modules are the same as those of the
main unit.
Transient operating conditions
The monitor is operated in normal use for a period not less than 20 minutes when moved from room
temperature (
20°C ± 2°C) to an environment of a temperature range from - 20 °C to 50 °C, and relative humidity
range from 15% to 95% (non-condensing).
The monitor is operated in normal use for a period not less than 20 minutes when moved from storage
temperature (range from - 30 °C to 70 °C) to room temperature (
20°C ± 2°C), and started up within 10 minutes
after the movement.
Components Property (unit) Operating Condition Storage Condition
Main Unit/
Transport Dock/AC
Adapter
Temperature (°C) 0 to 40 -30 to 70
Relative humidity
(noncondensing) (%)
5 to 95 5 to 95
Barometric (mmHg) 427.5 to 805.5 375 to 805.5 (with CO
2
)
120 to 805.5 (without CO
2
)
Modular Rack/
Dock
Temperature (°C) 0 to 40 -20 to 60
Relative humidity
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 427.5 to 805.5 120 to 805.5
Microstream CO
2
module
Temperature (°C) 0 to 40 -20 to 60
Relative humidity
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 430 to 790 430 to 790
Sidestream CO
2
module
Temperature (°C) 5 to 40 -20 to 60
Relative humidity
(noncondensing) (%)
15 to 95 10 to 95
Barometric (mmHg) 430 to 790 375 to 805.5