20 - 2 BeneVision N1 Patient Monitor Operator’s Manual
â–  A patient is admitted.
â–  A patient is discharged.
â–  Patient category is changed.
To set the default configuration, follow this procedure:
1. Select the
Main Menu quick key → from the Configuration column select Manage → input the required
password
→ Select .
2. Select
Select Default Config.
3. Select
Load the Latest Config or Load Specified Config.
â—† When you select Load Specified Config, the restored configuration is subject to the patient category
(adult, pediatric or neonate). This configuration can be either factory configuration or a saved user
configuration. As an example, select
Default Adult Config and then select Factory Default or user
configuration(s).
â—† When you select Load the Latest Config, the latest configuration is loaded when the monitor is
started or a patient is admitted.
20.5 Saving Current Settings
Current settings can be saved as a user configuration. Up to 10 user configurations can be saved.
To save current settings, follow this procedure:
1. Select the
Main Menu quick key → from the Configuration column select Manage → input the required
password
→ Select .
2. Select
Save Current Settings.
3. Input the configuration name.
4. Select
OK to save current settings as a user configuration.
20.6 Deleting a Configuration
To delete a configuration, follow this procedure:
1. Select the
Main Menu quick key → from the Configuration column select Manage → input the required
password
→ Select .
2. Select
Delete Configuration.
3. Select the configuration you want to delete:
â—† In the Delete Configuration dialog, selecting Local tab shows the existing user configurations on the
monitor.
â—† In the Delete Configuration dialog, selecting USB Drive tab shows the existing user configurations
on the USB drive.
4. Select Delete.
5. Select
OK.
20.7 Transferring a Configuration
When installing several monitors with identical user configurations, it is not necessary to set each unit
separately. Use a USB drive to transfer the configuration from monitor to monitor.
20.7.1 Exporting a Configuration
To export the current monitor’s configuration, follow this procedure:
1. Connect the N1 to the Dock.
2. Connect the USB drive to the Dock’s USB port.
3. Select the
Main Menu quick key → from the Configuration column select Manage → input the required
password
→ Select .

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals