BeneVision N1 Patient Monitor Operator’s Manual 21 - 23
21.16.3 Setting the SpO
2
Tone Mode
The monitor adjusts the QRS tone (pitch tone) according to the SpO
2
values. To set the SpO
2
tone mode, follow
this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Other tab.
3. Set
SpO2 Tone to Mode 1 or Mode 2. The pitch changes more rapidly in Mode 2 than in Mode 1, as
saturation levels change.
The SpO
2
tone mode chart for SpO
2
between 40% to 100% is as follows:
• It is highly recommended to use the same SpO
2
tone mode for the same monitors in a single area to
prevent issues arising from inconsistent operation.
21.16.4 Selecting the Language
To set the user interface (UI) language, follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Other tab
3. Set
Language.
21.16.5 Switching On or Off a Parameter
If there is no Parameters On/Off tab in the Screen Setup dialog then setting parameter switches is password
protected. Follow this procedure to set parameter switches:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Other tab.
3. Select
Parameters On/Off.
4. Enable or disable desired parameters.
When a parameter is switched off, the monitor stops data acquisition and alarming for this measurement.