BeneVision N1 Patient Monitor Operator’s Manual C - 5
C.3 SpO
2
Default Settings
Zero RR Alarm switch On
Priority High, unadjustable
Alarm Outputs Off
Zero RR Delay Adult: 20 sec
Pediatric: 20 sec
Neonate: 15 sec
RR Source Auto
Resp Lead Adult: Auto
Pediatric: Auto
Neonate: II
Gain ×2
Speed 6.25 mm/s
Auto Threshold Detection On
Item Default Setting
Item Default Setting
SpO
2
Alarm switch On
High limit Adult: 100%
Pediatric: 100%
Neonate: 95%
Low limit 90%
Priority Med
Alarm Outputs Off
SpO
2
Desat Alarm switch On
Low limit 80%
Priority High
Alarm Outputs Off
Satsecond (for Nellcor SpO
2
)Off
NIBP Simul Off
Display PI (for Masimo SpO
2
)On
Speed 25 mm/s