BeneVision N1 Patient Monitor Operator’s Manual 24 - 3
Verify that the equipment meets the following requirements:
â– Environment and power supply specifications are met.
â– The monitor housing and display screen are free from cracks or other damage.
â– The power cord is not damaged and the insulation is in good condition.
â– Connectors, plugs, and cables are not damaged or kinked.
â– Power cord and patient cables are securely connected to the equipment and modules.
24.5.2 Performing Power-on Test
The monitor automatically performs a power-on test at startup. Verify the following items during the power-on
test:
â– The equipment powers on properly.
â– The alarm system works properly, denoted by alarm sounds and display of all 3 alarm light colors.
â– The monitor displays properly.
24.5.3 Testing the Network Printer
To check the printer, follow this procedure:
1. Start a printing task to print waveforms and reports.
2. Check that the printer is properly connected and functions correctly.
3. Check that the printout is clear with no missing information.
24.5.4 Checking the Battery
For information on battery check, see 22.6.2Checking Battery Performance.
24.6 NIBP Maintenance
24.6.1 NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. The NIBP leakage test should be
performed once every two years or when you doubt the NIBP measurements. The NIBP leakage test should be
performed by Mindray-qualified service personnel only.
24.6.2 NIBP Accuracy Test
The NIBP accuracy test should be performed once every two years or when you doubt the NIBP measurements.
The NIBP accuracy test should be performed by Mindray-qualified service personnel only.
24.7 Disposing of the Monitor
Dispose of the monitor and its accessories when its service life is reached. Follow local regulations regarding the
disposal of such product.
• For disposal of parts and accessories, where not otherwise specified, follow local regulations
regarding disposal of hospital waste.