BeneVision N Series Patient Monitor Operator’s Manual E - 3
E.2 The Accuracy of Nellcor SpO
2
Sensors
SpO
2
Accuracy for Nellcor Sensors vs. Co-Oximeters (Arms)
M
odified Bland-Altman for SpO
2
- MAXAI, MAXPI, MAXII, MAXNI Sensors: (SpO
2
- SaO
2
)
vs.SaO
2
Modified Bland-Altman for SpO
2
- DS-100A Sensors: (SpO
2
- SaO
2
) vs.SaO
2
SpO
2
Range
100% to 70% 100% to 90% 90% to 80% 80% to 70%
DS-100A 1.64% 1.16% 1.67% 2.25%
D-YS, OXI-P/I, OXI-A/N
2.41% 1.38% 2.50% 3.60%
MAXAI, MAXPI, MAXII
1.62% 1.49% 1.57% 2.50%
MAXNI
1.85% 1.71% 1.51% 1.59%