BeneVision N1 Patient Monitor Operator’s Manual 21 - 1
21 Password-protected User Settings
This chapter describes the password-protected functions accessed in the Maintenance dialog.
21.1 Setting the Device Location
21.1.1 Setting Monitor Information
To set monitor information, follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Device Location tab.
3. Input monitor name, facility name and department name.
21.1.2 Setting Monitor Location
If your monitor is installed at a fixed position, or you do not need clinicians to change the room number and bed
number, you can set
Monitor Location to Fixed. To set the monitor location, follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Device Location tab.
3. Set
Location.
â—† Fixed: the Patient Management dialog displays Bed No. and Room No., but you cannot change
them.
â—† Unfixed: you can change Bed No. and Room No. from the Patient Management dialog.
4. Input the room number and bed number.
• If Location is set to Unfixed, Bed No. and Room No. are cleared from the Device Location tab each
time you discharge a patient.
21.2 Changing Patient Management Settings
21.2.1 Selecting Displayed Patient Information
You can define which items can be displayed and edited from the Patient Management tab. To do so, follow
this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Patient Management tab.
3. Select the fields you want to display in the
Patient Management dialog.
4. If necessary, select the customized fields and input names for these fields.
• If the monitor is connected with the CMS, the patient information items and customized fields are
loaded from the CMS.